Benefits from a Complete Product Solution
| Instrument |
|
Chemistry |
| |
Simultaneously tests up to 12 real time PCR samples |
|
Freeze-dried reagents; just add sample and water |
| |
Works in roughly 30 minutes |
|
Validated for sensitivity and specificity on the RAZOR EX System |
| Support |
Software |
| |
Obtain quick assistance through phone and online help desks |
|
Simple wizard-based user interface |
| |
Approved product for Homeland Security |
|
Automated qualitative analysis and text delineated exportable data |

User Interface
Internal calibration and diagnostics ensure the instrument is functioning within tolerances before and during operation. Automatic analysis software runs “on-the-fly,” with raw data and software analysis presented simultaneously.
- Designed for use in personal protective equipment.
- Large intuitive push-buttons
- Bright Color screen
- Bar code reader for protocol and pouch information uploads

Desktop Software
The software is an additional tool to expand the stand-alone RAZOR EX capabilities to aid in post-mission support and reachback.
Software features include:
- After incident data analysis
- Data archiving
- Report function
RAZOR EX System and Computer Specifications
| The RAZOR system can test for the following targets |
| Hazmat |
Anthrax, Brucella spp, Botulism A, Coxiella, E. coli 0157, Tularemia, Ricin, Salmonella, Smallpox, and Plague |
| Food Safety |
Campylobacter, L. monocytogenes, and Salmonella |
| Water Safety |
Cryptosporidium Target 1, E. coli 0157, and Salmonella |
| Instrument Specifications |
| Size (w x d x h) |
10" (25.4 cm) x 4.5" (11.4 cm) x 7.5" (19 cm) |
| Weight |
11 lb / 4.9 kg |
| Instrument Power Supply |
Rechargeable battery pack included with instrument |
| External Power Supply |
24 V DC, 4 A Output, 110–220 V AC |
| Performance Parameters |
Instrument set-up time ≤ 5 minutes
Sample run time ≤ 30 minutes (average) |
| Temperature Control |
Lower Heater: 56 °C to 64 °C / Upper Heater: 86 °C to 94 °C
Uniformity: 0.5 °C in single sample–Intersample / 1.0 °C across sample area–Intersample |
| Fluorescence Acquisition |
One-color optics modules
Excitation wavelength 470 ± 20 nm
Emission wavelength 530 ± 10 nm–TaqMan®
Emission wavelength 690 ± 60 nm–HybProbe™ |
| Software and Training |
| RAZOR EX Software |
Windows® XP application, downloads data for real-time offline analysis, intuitive navigation and after action reports. |
| Training |
The system comes with a training DVD with all the necessary multimedia tools for sampling, pouch loading, machine operation, maintenance, and software. |
| Ordering Information |
| |
Catalog No. |
RAZOR EX Instrument (690 Filter Set HP, Commercial Reagents)
|
RAZR-ASY-4000 |
RAZOR EX Instrument (510 Filter Set TQ, Hydrolysis Probe Reagents)
|
RAZR-ASY-4001 |
|
|
|
| Free Online Webinar |
BioDetection:
Issues & Answers

Learn from the those who have been in the hot zone dealing with critical issues relating to Biodetection.
Presented by David Ladd, Director, Hazardous Materials Counter-Terrorism Emergency Response Division, Massachusetts Department of Fire Services
Provided by:
 |
| Featured Accessory |

The LiNK2© is an easy-to-use, convenient and disposable no-power device that collects and purifies biological environmental samples in the field. It is designed to collect liquid and dry biological samples and quickly and conveniently prepares them for PCR analysis.
>> Click here for details
Distributed and
sold through
Triarii Scientific LLC.
|
|